WITTING INNOVATION LLC — Department of Health and Human Services SBIR Phase I: DESCRIPTION (provided by applicant): Capnometers can fail in a number of ways that are not

WITTING INNOVATION LLC — SBIR Phase I award from Department of Health and Human Services.

Amount
$149,322
Agency
Department of Health and Human Services · National Institutes of Health
Program / Phase
SBIR · Phase I
Solicitation
PA13-234
NAICS
Place of performance
UT
Period

Description

DESCRIPTION (provided by applicant): Capnometers can fail in a number of ways that are not detectable using current static calibration testing procedures. These failures could include leaks in the sample line, a weak sample pump, inadequate dynamic response, flow restriction and flow obstruction. Failure of a capnometer during critical care procedures such as sedation and anesthesia can be life threatening to patients. We plan to develop a patient breath simulator that will allow for the testing of capnometers under dynamic conditions seen during patient use. We propose to develop a portable device and control algorithms that will be capable of simulating CO2 concentrations and pressure waveforms mimicking human respiratory cycles. The same device will haveadditional testing modes that will identify capnometer failures that cannot be detected under static calibration. We propose to verify the ability of the device to detect real world failure modes by testing it on a number of capnometers that have been